The following steps must be carried out in the menu ´TIMES-CYCLES´ to
perform a
data recording in the true sleep mode:
1. Clear the measuring cycle:
MeasCycle: 00:00:00
2. Switch off the continuous measuring point scan:
ConvRate: 10 Cont: -
3.
Enter a print cycle with a min. duration of 2 minutes:
PrintCycle: 00:05:00
4.
Switch on the storage activation in the print cycle:
Store:
Sleep:-
"
5. Select the sleep mode:
Store:
"
Sleep:
-
6. Switch on the sleep mode using the key:
<
ON
>
"
7. In a meas. menu, start Measure by pressing key :
<
START
>
The device switches off, for control purposes
lamp ´SLEEP´ flashes
only the lamp ´
SLEEP
´ flashes regularly.
The start or stop through start and end
time or through limit values is, on
principle, not possible in sleep mode and,
therefore, must be switched off!
8. Within the set cycle the device switches
automatically on, performs a measuring point
scan, and then switches off again.
9.
Terminating the sleep mode:
Use the sliding switch (1) to switch the device
ON - OFF - ON
off and then on again.
10.Terminate measurement by pressing the key:
<
STOP
>
If the requirements 1 to 4 are not met, on start of a measurement, the device
will not enter the sleep mode but only the switching off the display.
The
SLEEP
lamp flashes more quickly than when in “sleep“ mode.
SLEEP
lamp flashes.
8. Switching the display on again by using the key:
ESC
If one minute expires without any key
being pressed the display will switch off again!
9. Terminate measurement by pressing the key:
<
STOP
>
ALMEMO
®
2590-9
49
Data Memory