Cancelling the memory output using the key:
ESC
Interrupting the memory output using the key:
<
STOP
>
The following options are available after an output interruption:
Recall individual measured values:
<
MANU
>
Re-start the automatic output:
<
START
>
Cancelling the memory output using the key:
ESC
After the memory output the device returns to the menu ´Memory´.
The memory contents will be output with the same print format as with a printer
operation, including multiple printouts and different formats (see manual 6.6.1).
Print Format:
MEMORY: 03
(evtl. connector number)
NUMBER: 12-001
(if activated)
DATE: 12.03.00
List format
12:30:00 01: +0012.0 °C NiCr Designation
below each other
02:!+0008.8 °C NiCr Water
03:>+125.00 °C Ntc MotorOil
The connector number of an external memory will be printed after each
headline ´MEMORY´.
Clear Memory
Menu ´Memory´:
Select function ´MemoryFree´ (see 5.4):
No: -- Free:
kB
0217.
To clear the memory, first press the key:
PROG
the memory capacity is highlighted:
No: -- Free:
kB
then the key:
<
CMEM
>
Full mem. capacity is indicated as avail. mem. space:
No: -- Free:0507.5kB
Or cancel by using the key:
ESC
8.6 Sleep Mode
For long term monitoring with larger measuring cycles it is possible to operate
the measuring device in sleep mode. Within this power saving mode the device
will be switched off after each measuring point scan and will be automatically
switched on for the next measuring point scan after the cycle time has expired.
This procedure allows, with one full charged battery to perform approximately
7000 measuring point scans. At 15 minutes for one cycle this results in a total
measuring time of more than 70 days.
To save energy only the display will be switched off if the print cycle is shorter
than two minutes or if, additionally, the measuring cycle or the continuous
measuring point scan has been programmed. In this energy saving mode the
operating time of the device is extended to approximately 65 hours for one
battery pack.
48
ALMEMO
®
2590-9
Data Memory