If the cleared channel is also the selected measuring channel, the current
measured value will be immediately displayed after the clearing process.
Furthermore, the peak values are cleared at each start of a measurement, if
the device has been correspondingly configured (standard setting, see 10.8).
6.1.3 Output of Menu Functions
Each data menu, together with all displayed functions, can be output via
interface to a printer or computer (connection of peripheral devices, see
manual 5.2). If you have selected the standard display, pressing the key
<PRINT> leads to a printout of the following listing:
Print data menu:
<
>
Meas. point, meas. value, designation:
01: +0023.5 °C Temperature
MAXVAL: 01: +0020.0 °C
MINVAL: 01: -0010.0 °C
PRINTTIMER: 00:01:23
Total memory capacity, free mem. in kB
MEMORY: S0512.1 F0324.4 A
The listing of the individual functions is given in section 6.6.1.
6.2 Measuring Point Scans and Output
Measuring point scans allow measured values of all measuring points to be
acquired, stored (if necessary) and recorded through a printer or computer
(see manual 6.5). If continuous measuring point scan has been selected (the
letter ´C´ will be indicated at the first digit of the status line), the available
measured values, on command, will only be stored or output.
6.2.1 Single Output/Storage of all Measuring Points
Single manual measuring point scans for acquiring the momentary measured
values of all active measuring points can be triggered by the key <MANU> (see
manual 6.5.1.1). The time will be started if it had been set to zero. If the true
time of day has to be indicated, it must first be set (see 7.1). The output format
can be set in the function ´PRINT CYCLE´ (see 7.2, list of printouts in manual
section 6.6.1).
Single manual measuring point scan:
<
MANU
>
The following symbols will,
for a short period
, be indicated in the
status line
:
The start arrow will light up and then go off again
´
!
´
Lightening up in case of a data output via interface
´
COM
´
Indicated when meas. values are being stored (see 8.2),
´
REC
´
With each press of the key the measured values are equally processed with
the corresponding measuring time.
ALMEMO
®
2590-9
23
Displaying a Measuring Point