9. SENSOR PROGRAMMING
As all ALMEMO
®
instruments contain the whole sensor programming stored in
the ALMEMO
®
connector plug, the user, usually, does not need to perform any
programming. Only if, for example, sensor errors must be corrected or existing
sensors must be scaled or limit values need to be specified the comprehensive
programming options have to be used.
If the corresponding sensor connector is
connected all parameters of a channel can
be checked and entered or changed via
keypad in the menu ´Sensor
Programming´. It must be considered that
standard sensors are, by a locking mode,
protected against unintentional
modification and that the locking level
must first be reduced before desired
changes can be performed (see 9.2).
Output of the sensor programming of all
active measuring points (command P15
see man. 6.2.3) by using the key:
<
>
9.1 Selecting the Input Channel
To query or to program the parameters of a sensor, it is first necessary to
select the menu ´Sensor Programming´ and then to set the required input
channel by using the key
M
▲
or
M
▼
. For this, only connected sensors and
activated channels will be considered. To be able to activate new channels, the
key ´MALL´ must be used to enable the selection of
all
channels. The key
´MACT´ can be used to reduce the selection back to the
active
channels only.
The corresponding connector number will be displayed with each input
channel.
When changing the input channel, the selected meas. point will not
be changed, i.e. a measurement in progress will not be interrupted.
Menu ´
SensorProgramming
´
* SENSORPROGRAMM *
Indication of connector number and input channel:
Connector:0
Channel:00
Selecting the next input channel using the key:
<
M
▲
>
Selecting the previous input channel using the key:
<
M
▼
>
Enabling the selection of all available channels:
<
MALL
>
Reducing the selection to all active channels:
<
MACT
>
50
ALMEMO
®
2590-9
Sensor Programming
* SENSOR PROGRAMM *
Connector:0 Chann.:00
Comment: Temperature
Averag.Mode: -----
Locking: 5.
7LimValMax: 35.0°C
LimValMin: -------
5BaseValue: -------
Factor: -------
Exponent: 0
4Zero Point: -------
Slope: -------
2Dimension: °C
1Range: Diff 01-00
MALL PRINT
M
▲
M
▼