6.1 Display of a Measuring Point
Different from previous instruments, the continuous measuring point scan (e.g.
after a reinitialisation, see 3.4) is the standard default setting of the ALMEMO
2590-9, i.e. all meas. points are continuously acquired and the meas. values
can be retrieved at any time, even if they are related to other channels (e.g.
with function channels, temperature or atmospheric pressure compensation).
6.1.1 Selecting the Measuring Point
The key <
M
▲
> allows for successively selecting all active meas. points
including the display of the current meas. value. If the key <
M
▲
> is pressed
and held (approx. 1s) the previous channel is again indicated. By selecting the
meas. channel the input channel is, at the same time, also selected (see 9.1).
Increasing the meas. channel using the key:
<
M
▲
>
Decreasing the meas. channel using the key:
<
M
▲
> press and hold
The measuring points, as well, can very easily be
increased or decreased by turning the
operating dial
.
6.1.2 Peak Value Memory with Time and Date
The highest and lowest value, including the
time and date, will each time be determined
and stored from the acquired measured
values of each measuring point. The
functions listed below are available to
display these values.
The software AMR-Control can be used to
easily load or configure the menu
´Monitoring´
with Max/Min times (illustrated
on the right) as a user menu (see 6.6).
Function max value:
MaxVal: 245.7°C
Function min value:
MinVal: 224.1°C
Function time and date of max value:
MaxTime: 12:34 01.02.
Function time and date of min value:
MinTime: 12:56 01.02.
To clear, select function (see 5.4):
MaxVal:
°C
245.7
Clear single value:
<
CLR
>
Clear max, min and average values of all channels:
<
CLRA
>
▲
▼
M
M
22
ALMEMO
®
2590-9
Displaying a Measuring Point
C
!
REC COM
|> >|
R01
"
12:34:56 01.02.00
PrintTimer:00:00:00nS
01
:
1254
.
5
°C
NiCr Temperature H
!
LimValMax: 250.0°C
Max Value: 245.7°C
Max Time:12:34 01.02.
LimValMin: 230.0°C
Min Value:
▼
224.1°C
Min Time:12:56 01.02.
M
▲
PRINT START MANU