5.2 Function Keys
The function of the keys F1 to F4 can be
different in the various menus. It is
indicated by abbreviations in the bottom
line of the display (soft keys). In the
manual the soft key abbreviations are set
in angle brackets, e.g. <START>.
Next to the meas. value, control symbols
for the meas.value are displayed
(s. below)
.
The following keys are available in the
standard display:
Selection of measuring points
F1
<
M
▲
>
Output of menu functions via the interface
F2
<
>
Starting a cyclic measurement
F3
<
START
>
Stopping a cyclic measurement
F3
<
STOP
>
Single manual output/storage of all measured values
F4
<
MANU
>
Returning back to the main menu
ESC
5.3 Control Symbols
Symbols for monitoring process control and battery in the status line:
Continuous measuring point scan:
C
Measurement stopped or started:
"
or
!
Measuring point scan started including storing:
REC
Measuring point scan started incl. output via interface:
COM
Start/end time of measurement is programmed:
|>
or
>|
Relay status (external output module) off or on:
R--
or
R01
Display illumination switched on:
#
Batt. charge state:100% charged,discharged
(connect charger)
:
,
flashes
Charger connected:
Symbols for checking the measured value (see above)
No sensor, measuring point deactivated:
´
-----
´
Measured value altered by sensor correction or scaling:
!
Averaging in progress:
M
¯
Output function altered (see 9.10.5):
D, H, L, M, A
Exceeding of limit value Max or Min:
▲
or
▼
flashes
Exceeding of measuring range: display of max. value
O
flashes
Undershooting of meas. range: display of min. value
U
flashes
Sensor breakage sensor voltage Lo : display
´
-.-.-
´
B
flashes
L
flashes
Supply voltage too low for measurement:
display switched off
ALMEMO
®
2590-9
17
C
!
REC COM
|> >|
R01
"
01
:
Temperature
°C
-
1254
.
5
Max Value: 245.7°C
Min Value: 224.1°C
PrintTimer: 00:00:13nU
MemoryFree: 0211.0kB
Display and Keypad
M
¯ H
!
▲
F1
F2
F3
F4
M
▲
PRINT START MANU