6.5.2 Menu List of Measured Values
The best overview of the meas. system
incl. all meas. points, time of day, date,
meas.cycle and print cycle is obtained with
the menu ´
MeasValList
´.
Remember to switch on the
continuous measuring point scan
(´C´) to ensure that the measured
values will be continuously updated.
This menu cannot be individually
configured, it can only be combined with
some selected functions:
At the first call a list with max meas. points is displayed
MeasValList: 20 MPts
It is not possible to display more than 20 channels.
00:
23.12°C
Functions can be allocated to a measured value by using:
<
FCT
>
This reduces the max. number of channels to 10.
The following function:
<
FCT
> .....
Measured value including
comment
:
MeasValList: Name
00:
23.12°C Temperature
Measured value including
max value
:
MeasValList: MaxVal
00:
23.12°C 32.67°C
Measured value including
min value
:
MeasValList: MinVal
00:
23.12°C 19.34°C
Measured value including
average value
:
MeasValList: AvgVal
00:
23.12°C 25.45°C
Measured value including
limit value Max
:
MeasValList: LV-Max
00:
23.12°C 32.67°C
Measured value including
limit value Min
:
MeasValList: LV-Min
00:
23.12°C 19.34°C
Measuring range
only (max. 20 channels):
MeasValList: Range
00:
NTC °C
Switching to the previous function:
press and hold <
FCT
>
Further channels can be displayed by
leafing through the list :
PROG
, <
M
▲
>, or <
M
▼
>.
The programming functions
can be accessed as follows :
PROG
, <
F
▲
/
▼
>,
PROG
.
ALMEMO
®
2590-9
35
C
!
REC COM
|> >|
R01
MeasValList: Name
12:34:56
01.02.00
M:
00:00:00
S P:
00:12:34
U
00:
23
.
12°C Temperat.
01:
▲
87.3°C NiCr
02:
123.4mV U2.4
10:
67.5%H Humidity
20:
15.2°C Dew Point
30:
11.2gk Mixture
FCT PRINT START MANU
Displaying Several Measuring Points