BenchCel Microplate Handling Workstation User Guide Addendum
12
Optimum Orientation
sensor threshold
Initial Orientation sensor offset
Measure A and B of your microplate.
Initial Orientation sensor offset = (A/2) + B = ___________ mm
Jog distance
Jog distance = Z
initial
– Z
adjusted
= _________ mm
Adjusted Orientation sensor offset
Adjusted Orientation sensor offset =
Initial Orientation sensor Jog distance = _______________ mm
Orientation
threshold
After you calculate the adjusted Orientation sensor offset, in BenchCel
Diagnostics, click
Move to Sensor
at each stack. In the
Notch Sensor
area,
the notches have lower readings than corners.
Highest Notch Sensor readings
Record the highest notch reading for each stack.
Highest reading among all stacks (Notch): ____________
Lowest Corner readings
Record the lowest corner reading for each stack.
Lowest reading among all stacks (Corner): ____________
Orientation threshold
Orientation threshold = Notch + ((Corner – Notch)/2) = __________
Stack 1
Stack 2
Stack 3
Stack 4
Stack 5
Stack 6
Stack 1
Stack 2
Stack 3
Stack 4
Stack 5
Stack 6
Содержание R Series
Страница 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division ...
Страница 10: ......
Страница 23: ......
Страница 24: ......
Страница 26: ......
Страница 34: ...Preface BenchCel Microplate Handling Workstation User Guide viii ...
Страница 60: ...Chapter 1 Introduction BenchCel Microplate Handling Workstation R Series User Guide 26 ...
Страница 124: ...Chapter 4 Performing a run BenchCel Microplate Handling Workstation R Series User Guide 90 ...
Страница 188: ...Appendix A BenchCel ActiveX control BenchCel Microplate Handling Workstation R Series User Guide 154 ...
Страница 210: ...Appendix B Quick reference BenchCel User Guide 176 ...
Страница 218: ...Index BenchCel Microplate Handling Workstation R Series User Guide 184 ...
Страница 219: ......
Страница 220: ...Agilent Technologies User Guide G5400 90001 ...