Chapter 5: Maintenance and troubleshooting
BenchCel Microplate Handling Workstation R-Series User Guide
122
2.
In the
Stack Settings
area, type a new value for the parameter you
want to change.
3.
Select
Home robot after protocol run finishes
if you want the robot
to return to its home position after the protocol run is finished.
4.
Optional
. To change the speed, click
Speed Settings.
See “Changing
the robot speed” on page 108 for instructions.
5.
When you are finished making the changes, click
OK
to apply and
save the new setting values.
Setting
Comment
Plate presence threshold
The value above which a plate is
considered present at the plate-
presence sensor. This value is
applied to all the sensors in all the
stacks.
Rack sensor threshold
The value above which a rack is
considered present at the rack-
presence sensor.
Low pressure threshold (psi)
The value at or below which the
system will display a low-pressure
warning.
Note: If you set this value too low,
the device could drop microplates
without generating a low air-
pressure warning. Velocity11
recommends that you set the
threshold at 20 psi.
Содержание R Series
Страница 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division ...
Страница 10: ......
Страница 23: ......
Страница 24: ......
Страница 26: ......
Страница 34: ...Preface BenchCel Microplate Handling Workstation User Guide viii ...
Страница 60: ...Chapter 1 Introduction BenchCel Microplate Handling Workstation R Series User Guide 26 ...
Страница 124: ...Chapter 4 Performing a run BenchCel Microplate Handling Workstation R Series User Guide 90 ...
Страница 188: ...Appendix A BenchCel ActiveX control BenchCel Microplate Handling Workstation R Series User Guide 154 ...
Страница 210: ...Appendix B Quick reference BenchCel User Guide 176 ...
Страница 218: ...Index BenchCel Microplate Handling Workstation R Series User Guide 184 ...
Страница 219: ......
Страница 220: ...Agilent Technologies User Guide G5400 90001 ...