Chapter 2: Unpacking and installing the workstation
BenchCel Microplate Handling Workstation R-Series User Guide
42
Checking the air
connections
To check the air connections:
1.
With the air source turned off, gently tug the air tubing at each
connection.
If you feel resistance at the connection, the tubing has been properly
installed.
2.
Turn on the air at the source (house, cylinder, or pump).
3.
Listen near each connection for hissing sounds that might indicate a
leak.
If you hear hissing sounds, turn off the air at the source, check and
tighten the connections, and then turn on the air again. If the
problem persists, contact your facilities department or Velocity11
Technical Support.
Disconnecting the
air source
You might need to disconnect the air tubing before moving or shipping
the BenchCel device and before performing maintenance or cleaning.
!! DAMAGE HAZARD !!
Do not pull the tubing out of the orange
quick disconnect fitting. Doing so can damage the fitting.
To disconnect the air tubing from the BenchCel device:
1.
Turn off the air at the source (house, cylinder, or pump).
2.
Push and hold the locking collar against the fitting, and then gently
pull the air tubing out.
Note:
Alternatively, you can use the SMC Pneumatics tool (TG-2)
to aid in this task. See the manufacturer’s documentation for use
instructions. Contact your local SMC parts supplier for ordering
details.
The following diagram shows a close-up view of the quick
disconnect fitting.
Push and hold the locking collar.
Gently pull out the tubing.
Содержание R Series
Страница 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division ...
Страница 10: ......
Страница 23: ......
Страница 24: ......
Страница 26: ......
Страница 34: ...Preface BenchCel Microplate Handling Workstation User Guide viii ...
Страница 60: ...Chapter 1 Introduction BenchCel Microplate Handling Workstation R Series User Guide 26 ...
Страница 124: ...Chapter 4 Performing a run BenchCel Microplate Handling Workstation R Series User Guide 90 ...
Страница 188: ...Appendix A BenchCel ActiveX control BenchCel Microplate Handling Workstation R Series User Guide 154 ...
Страница 210: ...Appendix B Quick reference BenchCel User Guide 176 ...
Страница 218: ...Index BenchCel Microplate Handling Workstation R Series User Guide 184 ...
Страница 219: ......
Страница 220: ...Agilent Technologies User Guide G5400 90001 ...