Chapter 3: Setting up the BenchCel Workstation
BenchCel Microplate Handling Workstation R-Series User Guide
70
8.
In the
BenchCel Diagnostics Controls
tab, click the
Jog/Teach
tab.
9.
Click
Go Home.
The robot head moves to its home position at the
center of the BenchCel Workstation.
10. In the
Motor Enabled
area, clear the check boxes for
X
and
Theta
to
disable the X and Theta motors.
The Z motor is still enabled (the Z
box is selected).
11. With the robot grippers holding the spare microplate, gently push or
pull the robot head and arms close to the teachpoint location.
Position the robot head and arms so that the microplate rests in the
target microplate location.