BenchCel Microplate Handling Workstation User Guide Addendum
6
4. Click
Move to Sensor.
The robot moves the first microplate into the
line of notch sensors.
5. Click the
Jog/Teach
tab and check the
Notch Sensor
readings. The
notches should have much lower readings than the corners. In
addition, the difference between the notch readings and corner
readings should be at a maximum.
In the following example, the Notch Sensor values indicate that the
microplate has two notches (3 and 7) and two corners (76 and 63).
6. To find the maximum difference between the notch readings and
corner readings:
a. Record the initial
z
-axis position (Z
initial
) in the upper left corner
of the BenchCel Diagnostics screen.
b. Jog the robot up or down, and then check the
Notch Sensor
readings. You can repeat this step until you find the maximum
difference between the notch readings and the corner readings.
c.
Record the adjusted
z
-axis position (Z
adjusted
).
Содержание R Series
Страница 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division ...
Страница 10: ......
Страница 23: ......
Страница 24: ......
Страница 26: ......
Страница 34: ...Preface BenchCel Microplate Handling Workstation User Guide viii ...
Страница 60: ...Chapter 1 Introduction BenchCel Microplate Handling Workstation R Series User Guide 26 ...
Страница 124: ...Chapter 4 Performing a run BenchCel Microplate Handling Workstation R Series User Guide 90 ...
Страница 188: ...Appendix A BenchCel ActiveX control BenchCel Microplate Handling Workstation R Series User Guide 154 ...
Страница 210: ...Appendix B Quick reference BenchCel User Guide 176 ...
Страница 218: ...Index BenchCel Microplate Handling Workstation R Series User Guide 184 ...
Страница 219: ......
Страница 220: ...Agilent Technologies User Guide G5400 90001 ...