BenchCel Microplate Handling Workstation User Guide Addendum
Worksheet for setting sensor thresholds
Plate-presence
threshold
Initial Plate present readings
Set the thresholds as follows:
❑
Plate presence threshold = 225 units
❑
Intensity = 100%,
Record the
Plate present
reading for each stack.
Plate present readings when microplates are unloaded
After adjusting the Intensity so that the highest Plate present reading is
less than or equal to 175, record the
Plate present
readings for each
stack.
Highest reading among all the stacks (P
unloaded
): _________________
Plate present readings when the microplates are loaded
After loading the microplates at each stack (
Load All
in BenchCel
Diagnostics), record the
Plate present
readings for each stack.
Lowest reading among all the stacks (P
loaded
): _________________
Plate presence threshold
Plate presence threshold = P
unloaded
+ ((P
loaded
– P
unloaded
)/2) = ______
Stack 1
Stack 2
Stack 3
Stack 4
Stack 5
Stack 6
Stack 1
Stack 2
Stack 3
Stack 4
Stack 5
Stack 6
Stack 1
Stack 2
Stack 3
Stack 4
Stack 5
Stack 6
Содержание R Series
Страница 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division ...
Страница 10: ......
Страница 23: ......
Страница 24: ......
Страница 26: ......
Страница 34: ...Preface BenchCel Microplate Handling Workstation User Guide viii ...
Страница 60: ...Chapter 1 Introduction BenchCel Microplate Handling Workstation R Series User Guide 26 ...
Страница 124: ...Chapter 4 Performing a run BenchCel Microplate Handling Workstation R Series User Guide 90 ...
Страница 188: ...Appendix A BenchCel ActiveX control BenchCel Microplate Handling Workstation R Series User Guide 154 ...
Страница 210: ...Appendix B Quick reference BenchCel User Guide 176 ...
Страница 218: ...Index BenchCel Microplate Handling Workstation R Series User Guide 184 ...
Страница 219: ......
Страница 220: ...Agilent Technologies User Guide G5400 90001 ...