Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
71
Using the Fluorescence Detector
3
Special Settings
Baseline Behavior
Baseline behavior allows you to choose what happens to
the baseline after a change in wavelength or PMT gain.
•
Append: Resets the baseline to the previous baseline
position.
•
Free: Makes no adjustment to the baseline position; you
may see a shift in baseline position.
•
Zero: Resets the baseline to the baseline position at the
start of the run. Note that this may not be 0 LU
(Luminescence Units), depending on the analysis
conditions.
Signal Polarity
Select the Signal Polarity you expect from your data from
Negative or Positive.