![Agilent Technologies G7121A Скачать руководство пользователя страница 27](http://html.mh-extra.com/html/agilent-technologies/g7121a/g7121a_user-manual_2867752027.webp)
Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
27
Introduction to the Fluorescence Detector
1
Analytical Information From Primary Data
Analytical Information From Primary Data
We now know how the primary data from your sample is acquired in the
optical unit. But how can the data be used as information in analytical
chemistry? Depending on the chemistry of your application, the luminescence
measured by the fluorescence detector will have different characteristics. You
must decide, using your knowledge of the sample, what mode of detection you
will use.
Fluorescence Detection
When the lamp flashes, the fluorescing compounds in your sample will
luminesce almost simultaneously, see
on page 27. The luminescence
is short-lived, therefore the fluorescence detector need only measure over a
short period of time after the lamp has flashed.
Figure 13
Measurement of Fluorescence
7ROG
7LPH>VHF@
,QWHQVLW\
,JQLWH
Содержание G7121A