Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
111
Optimizing the Detector
5
Example: Optimization for Multiple Compounds
4
Using the settings and the timetable (from previous page), do a second run
for the evaluation of the optimal excitation wavelength. See
Table 15
Peak #
Time
Emission Wavelength
1
5.3 min
330 nm
2
7.2 min
330 nm
3
7.6 min
310 nm
4
8.6 min
360 nm
5
10.6 min
445 nm
6
11.23 min
385 nm
DO NOT select
additional excitation
wavelengths (B, C,
D).
Doing so will
increase the scan
time and will lower
the performance.
Figure 40
Detector Settings for Excitation Scan