![Agilent Technologies G7121A Скачать руководство пользователя страница 44](http://html.mh-extra.com/html/agilent-technologies/g7121a/g7121a_user-manual_2867752044.webp)
44
Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
2
Site Requirements and Specifications
Site Requirements
Bench Space
The module dimensions and weight (see
on page 45) allow you to place
the module on almost any desk or laboratory bench. It needs an additional
2.5 cm (1.0 inches) of space on either side and approximately 8 cm
(3.1 inches) in the rear for air circulation and electric connections.
If the bench shall carry a complete HPLC system, make sure that the bench is
designed to bear the weight of all modules.
The module should be operated in a horizontal position.
Condensation
N O T E
Agilent recommends that you install the HPLC instrument in the InfinityLab Flex Bench
rack. This option helps to save bench space as all modules can be placed into one single
stack. It also allows to easily relocate the instrument to another Lab.
C A U T I O N
Condensation within the module
Condensation can damage the system electronics.
➔
Do not store, ship or use your module under conditions where temperature
fluctuations could cause condensation within the module.
➔
If your module was shipped in cold weather, leave it in its box and allow it to warm
slowly to room temperature to avoid condensation.
Содержание G7121A
Страница 86: ...86 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 4 Preparing the Module Solvent Information ...
Страница 230: ...230 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 10 Parts for Maintenance and Repair Kits ...
Страница 264: ...264 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 12 Hardware Information Early Maintenance Feedback ...
Страница 281: ...Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 281 Index ...