Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
103
Optimizing the Detector
5
Method Development
The two upper traces
are obtained using
two different
excitation
wavelengths.
The lower trace is a
pure standard of the
known impurities.
Figure 35
Qualitive analysis of MBC (2-benzimidazole carbamic acid methylester) and
impurities
SKHQD]LQH
GLDPLQRSKHQD]LQH
6WDQGDUG
7LPH>PLQ@
8QNQRZQ
Table 11
Conditions for analysis of DAP and MBC according to figures above
Column
Zorbax SB, 2 x 50 mm, PNA, 5 µm
Mobile phase
A = water; B = acetonitrile
Gradient
0 minutes, 5%
10 minutes, 15%
Flow rate
0.4 ml/min
Column temperature
35 °C
Injection volume
5 µl
FLD settings
PMT 12,
response time 4 s,
step size 5 nm
Ex 265 nm and 430 nm
Em 540 nm