Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
7
Error Information
Detector Error Messages
Detector Error Messages
Lamp Cover Open
Error ID: 6622, 6731
The lamp cover in the optical compartment has been removed. The lamp
cannot be turned on while this message is on.
FLF Board not found
Error ID: 6620, 6730
The FLF board could not be found by the main board (FLM). This message
comes together with some other message generated on the FLF board (e.g.
Leak, ...).
Probable cause
Suggested actions
1
Lamp cover removed.
Please contact your Agilent service
representative.
Probable cause
Suggested actions
1
FLF board not connected to the FLM board.
Please contact your Agilent service
representative.
2
Defective FLF board.
Please contact your Agilent service
representative.
3
Defective FLM board.
Please contact your Agilent service
representative.