Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
9
Maintenance
Flow Cell Flushing
Flow Cell Flushing
1
Flush with bidistilled water.
2
Flush with nitric acid (65 %) using a glass syringe.
3
Leave this solution in the cell for about one hour.
4
Flush thoroughly with bidistilled water at a higher flow rate, e.g.
1.5 mL/min.
When
If flow cell is contaminated
Tools required
Description
Glass syringe
Adapter
Parts required
#
Description
1
Bidistilled water, nitric acid (65 %), tubings to waste
WA R N I N G
Dangerous concentration of nitric acid
The nitric acid flushing procedure is not an infallible remedy for a dirty cell. It is to
be used as a last attempt to salvage the cell before cell replacement. Note that the
cell is a consumable item.
➔
Give proper attention to safety.
N O T E
Aqueous solvents in the flow cell can built up algae. Algae do fluoresce. Therefore do not
leave aqueous solvents in the flow cell for longer periods. Add a small percentage of
organic solvents (e.g. Acetonitrile or Methanol ~5 %).
N O T E
Do not exceed the pressure limit of 20 bar (0.2 MPa).
N O T E
Do not use the HPLC pump to pump nitric acid through the system.
Содержание G7121A
Страница 86: ...86 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 4 Preparing the Module Solvent Information ...
Страница 230: ...230 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 10 Parts for Maintenance and Repair Kits ...
Страница 264: ...264 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 12 Hardware Information Early Maintenance Feedback ...
Страница 281: ...Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 281 Index ...