Table 22
Raman Signal-to-Noise Test Conditions
Duration
approximately 23 minutes
Standard Flow Cell
G1321-60005, G5615-60005
Solvent
LC grade water, degassed
Flow rate
0.25 mL/min
Specification (single wavelength at signal)
>500 (according to settings in
Specification (single wavelength at
background)
>3000 (according to settings in
Specification (dual wavelength)
>300 (according to settings in
Table 23
Settings for Single Wavelength Specifications (at signal)
Time
EX
EM
PMT
Baseline
0
350
397
12
Free
20.30
350
450
12
Free
Table 24
Settings for Single Wavelength Specifications (at background)
Time
EX
EM
PMT
Baseline
0
350
450
14
Free
20.30
350
397
14
Free
Table 25
Settings for Dual Wavelength Specifications (Multi-EM Scan)
Time
EX
EM_A
EM_B
Spectra
From
To
Step
PMT
Baseline
Fit
Spectra
00.00
350
397
450
None
280
450
10
12
Free
OFF
20.30
350
450
450
None
280
450
10
12
Free
OFF
Содержание G7121A
Страница 86: ...86 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 4 Preparing the Module Solvent Information ...
Страница 230: ...230 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 10 Parts for Maintenance and Repair Kits ...
Страница 264: ...264 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 12 Hardware Information Early Maintenance Feedback ...
Страница 281: ...Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 281 Index ...