Test Functions and Calibration
8
D/A Converter (DAC) Test
Figure 72
D/A Converter (DAC) Test – Example of Integrator Plot
Running the Test with Instant Pilot
The test can be started via the command line.
1
To start the test
TEST: DAC 1
Reply:
RA 00000 TEST:DAC 1
2
To stop the test
TEST:DAC 0
Reply:
RA 00000 TEST:DAC 0
Test Evaluation
The noise on the step should be less than 3 μV.
Probable cause
Suggested actions
1
Bad cable or grounding problem between
detector and external device.
Check or replace the cable.
2
Defective detector main board.
Please contact your Agilent service
representative.