Test Functions and Calibration
8
Wavelength Accuracy Test
Figure 65
Example of bad EX/EM maxima (no maximum found)
Interpretation of the Results
If the test fails, check for:
✔
correctly positioned flow cell,
✔
clean flow cell (flush with isopropanol and clean bi-distilled water),
✔
no air bubble(s) (check via fluorescence scan or visual check of
cell/cuvette),
✔
solvent inlet filter (may create air bubbles in flow cell).
✔
check optical path for contamination (service)
✔
check alignment of lamp / trigger pack assembly (service)
✔
perform a Wavelength Calibration