![Agilent Technologies G7121A Скачать руководство пользователя страница 36](http://html.mh-extra.com/html/agilent-technologies/g7121a/g7121a_user-manual_2867752036.webp)
36
Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
1
Introduction to the Fluorescence Detector
System Overview
Figure 21
Infinity II Two Stack Leak Waste Concept (bench installation)
The waste tube connected to the leak pan outlet on each of the bottom
instruments guides the solvent to a suitable waste container.
Leak Sensor
C A U T I O N
Solvent incompatibility
The solvent DMF (dimethyl formamide) leads to corrosion of the leak sensor. The
material of the leak sensor, PVDF (polyvinylidene fluoride), is incompatible with DMF.
➔
Do not use DMF.