Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
151
Error Information
7
General Error Messages
Lost CAN Partner
Error ID: 0071
During an analysis, the internal synchronization or communication between
one or more of the modules in the system has failed.
The system processors continually monitor the system configuration. If one or
more of the modules is no longer recognized as being connected to the system,
the error message is generated.
Leak
Error ID: 0064
A leak was detected in the module.
The signals from the two temperature sensors (leak sensor and
board-mounted temperature-compensation sensor) are used by the leak
algorithm to determine whether a leak is present. When a leak occurs, the leak
sensor is cooled by the solvent. This changes the resistance of the leak sensor
which is sensed by the leak-sensor circuit on the main board.
Probable cause
Suggested actions
1
CAN cable disconnected.
•
Ensure all the CAN cables are connected
correctly.
•
Ensure all CAN cables are installed
correctly.
2
Defective CAN cable.
Exchange the CAN cable.
3
Defective main board in another module.
Switch off the system. Restart the system, and
determine which module or modules are not
recognized by the system.
Probable cause
Suggested actions
1
Loose fittings.
Ensure all fittings are tight.
2
Broken capillary.
Exchange defective capillaries.
Содержание G7121A
Страница 86: ...86 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 4 Preparing the Module Solvent Information ...
Страница 230: ...230 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 10 Parts for Maintenance and Repair Kits ...
Страница 264: ...264 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 12 Hardware Information Early Maintenance Feedback ...
Страница 281: ...Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 281 Index ...