Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
121
Optimizing the Detector
5
Design Features Help Optimization
Design Features Help Optimization
The module has several features you can use to optimize detection:
Check Performance Before You Start
Before you start you should check that your detector is performing according
to the specifications published by Agilent Technologies.
Your normal LC grade solvents may give good results most of the time but our
experience shows that baseline noise can be higher with LC grade solvents
than with fluorescence grade solvents.
Flush your solvent delivery system for at least 15 minutes before checking
sensitivity. If your pump has multiple channels, you should also flush the
channels not in use.
PMTGAIN
Amplification factor
LAMP
Flash frequency
RESPONSETIME
Data reduction interval