Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
117
Optimizing the Detector
5
How to collect spectra with modes SPECTRA ALL IN PEAK and APEX SPECTRA ONLY
2
In the FLD’s setup-screen there are 2 fields to enter the PKWD
Peakwidth
(Responsetime)
and the THRS
Threshold
(visible when
Multi-EX
or
Multi-EM
is
selected). Defaults are: PKWD = 6 (0.2 min); THRS = 5.000 LU.
The selected values are fixed during the run.
THRS and PDPW influence the peak-triggered spectra-acquisition. You can
change THRS in the FLD’s setup-screen; PDPW can only be changed with the
Peakwidth
-field in the
Timetable
.