Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
173
Test Functions and Calibration
8
Raman ASTM Signal-to-Noise Test
Raman ASTM Signal-to-Noise Test
This test verifies the Raman ASTM signal-to-noise for the FLD detectors.
Figure 59
Raman ASTM Signal-to-Noise Test (Lab Advisor)
Depending on the version of the detector the specification has changed.
Conditions: Standard flow cell (G1321-60005, G5615-60005), flow of
0.25 mL/min of water.
Table 21
Raman ASTM Signal-to-Noise Specification
Instrument
SNR Specification
Raman / Dark
SNR Specification
Dual WL
G7121A
500 / 3000
G7121B
500 / 3000
300
N O T E
The
Dark
and
Dual WL
values are just additional specifications. Only the
Raman
value is
used for the standard instrument checkout.
N O T E
The specification single wavelength at signal can be measured with the Agilent Lab
Advisor. All others (not used for standard checkout) have to be set up manually with the
information from