Agilent InfinityLab LC Series Diode Array Detectors User Manual
151
8
Test Functions and Calibration
Dark Current Test
Test Failed
Probable cause
Suggested actions
1
Defective slit assembly (stray light).
•
Run the
•
Run the
2
Defective detector main board.
Please contact your Agilent service representa
-
tive.
3
Defective PDA/optical unit.
Please contact your Agilent service representa
-
tive.