8
Test Functions and Calibration
Self-Test
Self-Test
The self-test runs a series of individual tests (described on the next pages), and
evaluates the results automatically. The following tests are run:
•
Slit Test (G7117B only)
•
Dark Current Test
•
Intensity Test
•
Wavelength Verification Test
•
ASTM Noise Test, a simplified version of the ASTM Drift and Noise Test
(without testing the Drift)
When
For complete detector check.
Parts required
#
Description
1
Max-Light Cartridge Cell (filled with water)
OR
1
Max-Light Cartridge Test Cell
Preparations
•
Lamp must be on for at least 10 minutes.
•
For noise test a longer warm-up time may be required (> 2 hours).
•
When using a Max-Light Cartridge Cell a flow rate of 0.5 mL/min with water is required.