Agilent InfinityLab LC Series Diode Array Detectors User Manual
140
8
Test Functions and Calibration
Slit Test
Slit Test
Slit Test (G7117B)
The slit test verifies correct operation of the micromechanical slit.
During the test, the slit is moved through all slit positions while the detector
monitors the lamp intensity change. When the slit position is changed, the
intensity drop (move to smaller slit) or intensity increase (move to larger slit)
must be within a defined range.
If the intensity changes are outside the expected range, the test fails.
1
Run the
Slit Test
with the Agilent Lab Advisor (for further information see
Online-Help of user interface).
Figure 43 Slit Test – Results
When
In case of problems.
Parts required
#
Description
1
Max-Light Cartridge Cell (filled with water)
OR
1
Max-Light Cartridge Test Cell
Preparations
•
Lamp must be on for at least 10 minutes.
•
When using a Max-Light Cartridge Cell a flow rate of 0.5 mL/min with water is required.