![Agilent Technologies G7117C Скачать руководство пользователя страница 157](http://html1.mh-extra.com/html/agilent-technologies/g7117c/g7117c_user-manual_2867751157.webp)
Agilent InfinityLab LC Series Diode Array Detectors User Manual
157
9
Maintenance
Cleaning the Module
Cleaning the Module
To keep the module case clean, use a soft cloth slightly dampened with water, or
a solution of water and mild detergent. Avoid using organic solvents for cleaning
purposes. They can cause damage to plastic parts.
Remove and Install Doors
WARNING
Liquid dripping into the electronic compartment of your module can cause
shock hazard and damage the module
Do not use an excessively damp cloth during cleaning.
Drain all solvent lines before opening any connections in the flow path.
NOTE
A solution of 70 % isopropanol and 30 % water might be used if the surface of the
module needs to be disinfected.
Parts required
p/n
Description
5067-5737
Door left
5067-5736
Door right
NOTE
The figures shown in this procedure exemplarily show the Infinity II Multisampler
module.
The principle of how to remove and/or install doors works in the same way for all
Infinity II modules.