Agilent InfinityLab LC Series Diode Array Detectors User Manual
163
9
Maintenance
Replace the Max-Light Cartridge Cell
Install the Max-Light Cartridge Cell
3
Flip the cartridge lever towards the front (down).
4
Pull the cartridge holder completely out.
5
Remove the flow cell cartridge.
1
Insert the flow cell cartridge.
2
Slide the cartridge holder completely into the detector.