Agilent InfinityLab LC Series Diode Array Detectors User Manual
200
12
Hardware Information
Setting the 6-bit Configuration Switch
Special Settings
Boot-Resident/Main
Firmware update procedures may require this mode in case of firmware loading
errors (main/resident firmware part).
If you use the following switch settings and power the instrument up again, the
instrument firmware stays in the resident/main mode. In resident mode, it is not
operable as a module. It only uses basic functions of the operating system for
example, for communication. In this mode the main firmware can be loaded
(using update utilities).
Table 30
6-bit Configuration Switch
Mode
Function/Setting
Switch 1
Switch 2
Switch 3
Switch 4
Switch 5
Switch 6
COM
1
1
When selecting mode COM, settings are stored to non-volatile memory. When selecting mode TEST, COM settings are taken from
non-volatile memory.
0
n.a.
2
2
not assigned - Always keep these switches on position ‘0’ (off)
n.a.
LAN Init Mode
n.a.
Use Default IP Address
3
3
Default IP Address is 192.168.254.11
0
0
0
0
0
Use Stored IP Address
0
0
0
1
0
Use DHCP to request IP Address
4
4
Host Name will be the MAC address.
0
0
1
0
0
Test
1
System
n.a.
n.a.
n.a.
ColdStart
Boot Main System / Keep Data
0
0
0
0
0
Boot Resident System / Keep Data
1
0
0
0
0
Boot Main System / Revert to
Default Data
0
0
0
0
1
Boot Resident System / Revert to
Default Data
1
0
0
0
1