Agilent InfinityLab LC Series Diode Array Detectors User Manual
204
12
Hardware Information
Early Maintenance Feedback
Setting the EMF Limits
The setting of the
EMF
limits must be optimized over one or two maintenance
cycles. Initially the default
EMF
limits should be set. When instrument
performance indicates maintenance is necessary, take note of the values
displayed by the
EMF counters
. Enter these values (or values slightly less than the
displayed values) as
EMF
limits, and then reset the
EMF counters
to zero. The
next time the
EMF counters
exceed the new
EMF
limits, the
EMF
flag will be
displayed, providing a reminder that maintenance needs to be scheduled.
NOTE
This function is only available via Agilent Lab Advisor or Instant Pilot.