![Agilent Technologies G7117C Скачать руководство пользователя страница 83](http://html1.mh-extra.com/html/agilent-technologies/g7117c/g7117c_user-manual_2867751083.webp)
Agilent InfinityLab LC Series Diode Array Detectors User Manual
83
5
Optimizing the Detector
Optimizing for Sensitivity, Selectivity, Linearity and Dispersion
Optimizing for Sensitivity, Selectivity, Linearity
and Dispersion
Flow Cell Path Length
Lambert-Beer’s law shows a linear relationship between the flow cell path length
and absorbance.
where
T is the transmission, defined as the quotient of the intensity of the transmitted
light I divided by the intensity of the incident light, I
0
,
is the extinction coefficient, which is a characteristic of a given substance under
a precisely-defined set of conditions of wavelength, solvent, temperature and
other parameters,
C [mol/L] is the concentration of the absorbing species, and
d [m] is the path length of the cell used for the measurement.
The detector can now output the signal in two forms:
1
In Absorbance divide by the path length AU/cm, that is then similar to [
x C].
Advantage: samples with same concentration have same peak height also at
cells with different path lengths.
The upper limit of concentration: the linearity limit of the detector is then seen
at about 2 AU/path length, so for the 6 cm Max-Light Cartridge Cell the
linearity limit is 333 mAU/cm].
2
In AU that is equal to
x C x d like normal done in the past: now for
recalculation to your concentration C the path length must be considered.
Therefore, flow cells with longer path lengths yield higher signals. Although noise
usually increases little with increasing path length, there is a gain in
signal-to-noise ratio.
When increasing the path length, the cell volume could increase. Depending on
the peak volume, this could cause more peak dispersion.