Agilent InfinityLab LC Series Diode Array Detectors User Manual
130
8
Test Functions and Calibration
Intensity Test
Intensity Test
The intensity test measures the intensity of the UV-lamp over the full wavelength
range (190 - 640 nm). Four spectral ranges are used to evaluate the intensity
spectrum. The test is used to determine the performance of the lamp and optics
on page 132). When the test is started, the 1-nm slit is
moved into the light path automatically (G7117B only). On the G7117A/G7117C,
the 4 nm fixed slit is used. To eliminate effects due to absorbing solvents, the
test should be done with water in the Max-Light Cartridge Cell or with the
Max-Light Cartridge Test Cell. The shape of the intensity spectrum is primarily
dependent on the lamp, grating, and diode array characteristics. Therefore,
intensity spectra will differ slightly between instruments.
1
Run the
Intensity-Test
with Agilent Lab Advisor (for further information see
Online-Help of user interface).
Figure 35 Intensity Test – Results
When
In case of UV-lamp problem (drift, noise).
Parts required
#
Description
1
Max-Light Cartridge Cell (filled with water)
OR
1
Max-Light Cartridge Test Cell
Preparations
Lamp must be on for at least 10 minutes.