Agilent InfinityLab LC Series Diode Array Detectors User Manual
168
9
Maintenance
Replace Leak Handling System Parts
Replace Leak Handling System Parts
Parts required
p/n
Description
5043-0856
Leak Adapter
5063-6527
Tubing, Silicon Rubber, 1.2 m, ID/OD 6/9 mm
approximately 85 mm required
1
Open the doors.
2
Locate the Leak Adapter (1) and Tubing (2).
3
Press the Leak Adapter down and remove it together
with the tubing.
4
Install the Leak Adapter by pressing it into the Main
Cover.
(1)
(2)