Agilent InfinityLab LC Series Diode Array Detectors User Manual
132
8
Test Functions and Calibration
Cell Test
Cell Test
The cell test measures the intensity of the UV-lamp over the full wavelength
range (190 - 690 nm), once with the Max-Light Cartridge Cell installed, and once
with the Max-Light Cartridge Test Cell. The resulting intensity ratio is a measure
of the amount of light absorbed by the Max-Light Cartridge flow cell. The test can
be used to check for dirty or contaminated flow cell windows. When the test is
started, the 1-nm slit is moved into the light path automatically (G7117B only). On
the G7117A/G7117C, the 4 nm fixed slit is used.
This test should be performed initially with a new detector/flow cell. The values
should be kept for later reference/comparison.
5
Old UV-lamp.
Exchange the UV-lamp.
6
Defect optical unit.
If the test fails with Max-Light Cartridge Test Cell
and new UV-lamp, please contact your Agilent
service representative.
Probable cause
Suggested actions
NOTE
If only one range fails and the application does not require this range, the lamp
may not be changed.
When
In case of low intensity or noise and drift problem.
Parts required
#
Description
1
Max-Light Cartridge Cell (filled with water)
1
Max-Light Cartridge Test Cell
Preparations
•
Lamp must be on for at least 10 minutes.
•
When using a Max-Light Cartridge Cell a flow rate of 0.5 mL/min with water is required.