Agilent InfinityLab LC Series Diode Array Detectors User Manual
32
2
Site Requirements and Specifications
Performance Specifications
Performance Specifications G7117B
Table 3
Performance Specifications G7117B
Type
Specification
Detection type
1024-element photodiode array
Light source
Deuterium lamp
Number of signals
8
Maximum data rate
240 Hz (both spectra and signals)
Short term signal
noise (ASTM)
With 10 mm Max-Light cartridge cell:
<±
3·10
-6
AU at 230 nm, slit width
4 nm, TC 2 s
With 60 mm Max-Light cartridge cell:
<±
0.6·10
-6
AU at 230 nm, slit width
4 nm, TC 2 s
Drift
<
0.5·10
-3
AU/h at 230 nm
Linear absorbance
range
>
2 AU (5 %) at 265 nm, typically 2.5 AU (5 %)
Wavelength range
190 – 640 nm
Wavelength accuracy
±
1 nm, self-calibration with deuterium lines
Wavelength precision
<±
0.1 nm
Wavelength bunching 2 – 400 nm, programmable in 1 nm steps
Slit width
1, 2, 4, 8 nm
Diode width
~0.5 nm
Nominal spectral reso
-
lution
< 1 nm
Time programmable
Wavelength, polarity, peak width, lamp bandwidth, auto balance, wavelength
range, threshold, spectra storage mode
Flow cells
User-exchangeable, self-aligning cartridge cells with RFID tags.
Max-Light Cartridge Cell (Standard): 10 mm,
V
= 1.0 µL
Max-Light Cartridge Cell (High Sensitivity): 60 mm,
V
= 4 µL
Max-Light Cartridge Ultra Low Dispersion (ULD) Cell: 10 mm,
V
= 0.6 µL
Max-Light Cartridge High Dynamic Range (HDR) Cell: 3.7 mm,
V
= 0.9 µL
Max-Light Cartridge Cell LSS: 10 mm,
V
= 1.0 µL
Maximum Operating Pressure (MOP)
1
: 70 bar (60 bar for LSS)
Maximum Incidental Pressure (MIP)
2
: 150 bar