Agilent InfinityLab LC Series Diode Array Detectors User Manual
126
8
Test Functions and Calibration
Use of Max-Light Cartridge Test Cell
Use of Max-Light Cartridge Test Cell
The Max-Light Cartridge Test Cell is recommended to be used for several tests
instead of the Max-Light Cartridge Cell (10 mm, V(
) = 1 µL) or the Max-Light
Cartridge Cell (60 mm, V(
) = 4 µL) because it allows running the test(s) without
any influence of the rest of the system (degasser, pump, sampler and others).
The results of the test cell are comparable with the Max-Light Cartridge Cell
(10 mm, V(
) = 1 µL) filled with water, e.g. Intensity Profile. Only the Absorbance
value is higher on the Max-Light Cartridge Cell.
If the profile of the Max-Light Cartridge Cell differs in the low UV range, then
absorbing solvents are in the cell and should be flushed out. See also
Below table gives an idea on the signal height variation of the Max-Light Cartridge
Cells compared to Max-Light Cartridge Test Cell.
NOTE
When using the Max-Light Cartridge Cell for tests/calibrations, it should be run at
0.5 mL/min constant flow with water. This assures that the light path is always
flushed.
Table 23
Max-Light Cartridge Cells compared to Max-Light Cartridge Test Cell
Part Number
Description
Signal Height (typical)
G4212-60011
Max-Light Cartridge Test Cell
100 %
G4212-60008
Max-Light Cartridge Cell 10 mmV(
) = 1 µL
~ 100 %
G4212-60007
Max-Light Cartridge Cell 60 mmV(
) = 4 µL
~ 100 %
G4212-60032
Max-Light Cartridge Cell HDR (3.7 mm, V(
) 0.4 µL)
100 %
G4212-60017
Max-Light Cartridge Cell ULD (10 mm, V(
) 0.6 µL)
100 %
G7117-60020
Max-Light Cartridge Cell LSS (10 mm, V(
) 1.0 µL)
~ 100 %