Agilent InfinityLab LC Series Diode Array Detectors User Manual
156
9
Maintenance
Overview of Maintenance
Overview of Maintenance
The following pages describe maintenance (simple repairs) of the detector that
can be carried out without opening the main cover.
Table 24
Overview of Maintenance
Procedure
Typical Frequency
Notes
Cleaning of module
If required
Deuterium lamp
exchange
If noise and/or drift exceeds your application limits or
lamp does not ignite.
A wavelength calibration test and an
intensity test should be performed after
replacement.
Flow cell exchange
If leaking or if intensity drops due to contaminated flow
cell.
A wavelength calibration test should be
performed after replacement.
Leak sensor drying
If leak has occurred.
Check for leaks.
Leak handling System
replacement
If broken or corroded.
Check for leaks.