CI Maintenance
6
5977 Series MSD Operation Manual
157
To Remove the CI Ion Source
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Pliers, long-nose (8710-1094)
Procedure
1
Vent the MSD. See
2
Open the analyzer chamber. See
“To Open the Analyzer Chamber”
Make sure you use an antistatic wrist strap and take other antistatic
precautions before touching analyzer components.
3
Disconnect the seven wires at the ion source. Use the pliers to pull on the
metal connectors at the source. Do not bend the wires any more than
necessary. See
158 for color coding of the wires,
Source feedthrough board
Ion source
Thumbscrews
Source heater and temperature
sensor wires
Source radiator