![Agilent Technologies G7035A Скачать руководство пользователя страница 93](http://html1.mh-extra.com/html/agilent-technologies/g7035a/g7035a_operation-manual_2868127093.webp)
Operating in Chemical Ionization (CI) Mode
4
5977 Series MSD Operation Manual
93
CI Autotune
After the reagent gas flow is adjusted, the lenses and electronics of the MSD
should be tuned. See
Perfluoro-5,8-dimethyl-3,6,9-trioxidodecane (PFDTD) is used as the calibrant.
Instead of flooding the entire vacuum chamber, the PFDTD is introduced
directly into the ionization chamber through the GC/MSD interface by means
of the gas flow control module.
There is a PCI autotune for methane only, as there are no PFDTD ions
produced by other gases in positive mode. PFDTD ions are visible in NCI for
any reagent gas. Always tune for methane PCI first, regardless of which mode
or reagent gas you wish to use for your analysis.
There are no tune performance criteria. If CI autotune completes, it passes.
EMVolts (electron multiplier voltage) at or above 2600
V, however, indicates a
problem. If your method requires EMVolts set at +400, you may not have
adequate sensitivity in your data acquisition.
C A U T I O N
After the source is changed from EI to CI or vented for any other reason, the MSD must
be purged and baked out for at least 2 hours before tuning. Longer bakeout is
recommended before running samples requiring optimal sensitivity.
C A U T I O N
Always verify MSD performance in EI before switching to CI operation. See
71. Always set up the CI MSD in PCI first, even if you are
going to run NCI.
Содержание G7035A
Страница 1: ...Agilent Technologies Agilent 5977 Series MSD Operation Manual ...
Страница 30: ...30 5977 Series MSD Operation Manual 1 Introduction ...
Страница 88: ...88 5977 Series MSD Operation Manual 3 Operating in Electron Ionization EI Mode ...
Страница 112: ...112 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 18 PCI autotune report ...
Страница 114: ...114 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 19 NCI autotune ...
Страница 170: ...170 5977 Series MSD Operation Manual 6 CI Maintenance ...
Страница 171: ......
Страница 172: ... G3870 90003 G3870 90003 Agilent Technologies Agilent Technologies Inc Printed in USA February 2013 ...