5977 Series MSD Operation Manual
1
Introduction
Dangers unique to GC/MSD operation
Hydrogen presents a number of dangers. Some are general, others are unique
to GC or GC/MSD operation. Dangers include, but are not limited to:
•
Combustion of leaking hydrogen
•
Combustion due to rapid expansion of hydrogen from a high-pressure
cylinder
•
Accumulation of hydrogen in the GC oven and subsequent combustion (see
your GC documentation and the label on the top edge of the GC oven door)
•
Accumulation of hydrogen in the MSD and subsequent combustion
Hydrogen accumulation in an MSD
All users should be aware of the mechanisms by which hydrogen can
accumulate (
21) and know what precautions to take if they
know or suspect that hydrogen has accumulated. Note that these mechanisms
apply to
all
mass spectrometers, including the MSD.
WA R N I N G
The MSD cannot detect leaks in inlet and/or detector gas streams. For this reason,
it is vital that column fittings should always be either connected to a column or have
a cap or plug installed.
Содержание G7035A
Страница 1: ...Agilent Technologies Agilent 5977 Series MSD Operation Manual ...
Страница 30: ...30 5977 Series MSD Operation Manual 1 Introduction ...
Страница 88: ...88 5977 Series MSD Operation Manual 3 Operating in Electron Ionization EI Mode ...
Страница 112: ...112 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 18 PCI autotune report ...
Страница 114: ...114 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 19 NCI autotune ...
Страница 170: ...170 5977 Series MSD Operation Manual 6 CI Maintenance ...
Страница 171: ......
Страница 172: ... G3870 90003 G3870 90003 Agilent Technologies Agilent Technologies Inc Printed in USA February 2013 ...