5977 Series MSD Operation Manual
5
General Maintenance
4
Remove the electron multiplier horn.
5
Install the new electron multiplier horn.
6
Close the retaining clip.
The signal pin on the horn must rest
on the outside
of the loop in the
contact strip.
Do not
put the signal pin on the inside of the loop in the
contact strip. Incorrect installation will result in poor sensitivity or no
signal.
7
Close the analyzer chamber. See
“To Close the Analyzer Chamber”
8
Pump down the MSD. See