100
5977 Series MSD Operation Manual
4
Operating in Chemical Ionization (CI) Mode
• Ion polarity
Always set up in PCI with methane first, then switch to your
desired ion polarity and reagent gas.
• Abundance target
Adjust higher or lower to get the desired signal
abundance. Higher signal abundance also gives higher noise abundance.
This is adjusted for data acquisition by setting the EMV in the method.
• Peakwidth target
Higher peakwidth values give better sensitivity, lower
values give better resolution.
• Maximum emission current
Optimum emission current maximum for NCI
is very compound-specific and must be selected empirically. Optimum
emission current for pesticides, for example, may be about 200 µA.