Operating in Electron Ionization (EI) Mode
3
5977 Series MSD Operation Manual
55
Before You Turn On the MSD
Verify the following
before
you turn on or attempt to operate the MSD.
•
The vent valve must be closed (the knob turned all the way clockwise).
•
All other vacuum seals and fittings must be in place and fastened correctly.
The front side plate screw should not be tightened, unless hazardous carrier
or reagent gasses are being used.
•
The MSD is connected to a grounded power source.
•
The GC/MSD interface extends into the GC
oven.
•
A conditioned capillary column is installed in the GC inlet and in the
GC/MSD interface.
•
The GC is on, but the heated zones for the GC/MSD interface, the GC inlet,
and the oven are off.
•
Carrier gas of at least 99.9995% purity is plumbed to the GC with the
recommended traps.
•
If hydrogen is used as carrier gas, carrier gas flow must be off and the front
sideplate thumbscrew must be loosely fastened.
•
The foreline pump exhaust is properly vented.
WA R N I N G
The exhaust from the foreline pump contains solvents and the chemicals you are
analyzing. If using the standard foreline pump, it also contains traces of pump oil. If
you are using toxic solvents or analyzing toxic chemicals, remove the oil trap
(standard pump) and install a hose (11-mm id) to take the foreline pump exhaust
outside or to a fume (exhaust) hood. Be sure to comply with local regulations. The
oil trap supplied with the standard pump stops only pump oil. It does not trap or filter
out toxic chemicals.
WA R N I N G
If you are using hydrogen as a carrier gas, do not start carrier gas flow until the
MSD has been pumped down. If the vacuum pumps are off, hydrogen will
accumulate in the MSD and an explosion may occur. Read
Содержание G7035A
Страница 1: ...Agilent Technologies Agilent 5977 Series MSD Operation Manual ...
Страница 30: ...30 5977 Series MSD Operation Manual 1 Introduction ...
Страница 88: ...88 5977 Series MSD Operation Manual 3 Operating in Electron Ionization EI Mode ...
Страница 112: ...112 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 18 PCI autotune report ...
Страница 114: ...114 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 19 NCI autotune ...
Страница 170: ...170 5977 Series MSD Operation Manual 6 CI Maintenance ...
Страница 171: ......
Страница 172: ... G3870 90003 G3870 90003 Agilent Technologies Agilent Technologies Inc Printed in USA February 2013 ...