CI Maintenance
6
5977 Series MSD Operation Manual
159
To Disassemble the CI Ion Source
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Hex ball driver, 1.5 mm (8710-1570)
•
Hex ball driver, 2.0 mm (8710-1804)
•
Wrench, open-end, 10 mm (8710-2353)
Procedure
1
Remove the ion source. See
2
Remove the filaments. Refer to
3
Separate the repeller assembly from the source body. The repeller
assembly includes the source heater block assembly, repeller, and related
parts.
4
Remove the repeller and ceramic insulator and separate them.
5
Remove the setscrew for the lenses.
6
Pull the lens assembly out of the source body.
7
Remove the drawout cylinder and drawout plate from the source body.
8
Separate the ion focus lens, entrance lens and insulator.