Installing GC Columns
2
5977 Series MSD Operation Manual
39
Procedure
1
Condition the column. (See
“To Condition a Capillary Column”
2
If you are not using Quick Swap, vent the MS. To vent the MS, see
3
Slide an interface nut and conditioned ferrule onto the free end of the GC
column. The tapered end of the ferrule must point towards the nut.
4
Use the column cutter to score the column 2
cm from its end.
5
While holding the column against the column cutter with your thumb,
break the column against the edge of the column cutter.
6
Inspect the end for jagged edges or burrs. If the break is not clean and
even, repeat steps
4 and 5.
7
Wipe the end with alcohol.
8
Insert the column into the column measuring tool.
9
Slide the column so that the end projects 1–2
mm past the end of the tool.
10
Finger tighten the fitting.
11
Tighten 1/4 to 1/2 turn to fix the ferrule to the column.
12
Slide the column into the GC/MS interface.
WA R N I N G
The analyzer, GC/MS interface, and other components in the analyzer chamber
operate at very high temperatures. Do not touch any part until you are sure it is cool.
WA R N I N G
Dangerous voltages exist inside the analyzer chamber, which can result in fatal
injury. Do not open the analyzer chamber door for any reason. If access is ever
required, trained service personnel must first disconnect the instrument from the
building power source.
WA R N I N G
The GC operates under high temperatures. To avoid burns, do not touch any GC parts
until you are certain they are cool.
Содержание G7035A
Страница 1: ...Agilent Technologies Agilent 5977 Series MSD Operation Manual ...
Страница 30: ...30 5977 Series MSD Operation Manual 1 Introduction ...
Страница 88: ...88 5977 Series MSD Operation Manual 3 Operating in Electron Ionization EI Mode ...
Страница 112: ...112 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 18 PCI autotune report ...
Страница 114: ...114 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 19 NCI autotune ...
Страница 170: ...170 5977 Series MSD Operation Manual 6 CI Maintenance ...
Страница 171: ......
Страница 172: ... G3870 90003 G3870 90003 Agilent Technologies Agilent Technologies Inc Printed in USA February 2013 ...